On the Criteria for Classifying Products as Medical Devices within the Eurasian Economic Union
Recommendation No. 25 of the Board of the Eurasian Economic Commission dated November 12, 2018.
The Board of the Eurasian Economic Commission, in accordance with Article 31 of the Treaty on the Eurasian Economic Union dated May 29, 2014, and paragraph 2 of Article 3 of the Agreement on Common Principles and Rules for the Circulation of Medical Devices (Medical Products and Medical Equipment) within the Eurasian Economic Union dated December 23, 2014,
for the purpose of eliminating differences in the requirements imposed when classifying products as medical devices,
recommends that the Member States of the Eurasian Economic Union, upon the expiration of 6 months from the date of publication of this Recommendation on the official website of the Eurasian Economic Union, apply the Criteria for Classifying Products as Medical Devices within the Eurasian Economic Union in accordance with the Appendix.
| Chairman of the Board of the Eurasian Economic Commission | T. Sargsyan |
|---|
| APPENDIX to Recommendation No. 25 of the Board of the Eurasian Economic Commission dated November 12, 2018 |
|---|
CRITERIA for Classifying Products as Medical Devices within the Eurasian Economic Union
I. General Provisions
- This document has been developed in accordance with Article 31 of the Treaty on the Eurasian Economic Union dated May 29, 2014, and paragraph 2 of Article 3 of the Agreement on Common Principles and Rules for the Circulation of Medical Devices (Medical Products and Medical Equipment) within the Eurasian Economic Union dated December 23, 2014, and establishes the basic principles for classifying products as medical devices within the Eurasian Economic Union (hereinafter referred to as the Union) in cases where this cannot be unambiguously established on the basis of the definitions established by the acts forming part of the law of the Union.
- This document may be used when preparing documents for the registration and examination of medical devices in accordance with the Rules for Registration and Examination of the Safety, Quality and Effectiveness of Medical Devices, approved by Decision No. 46 of the Council of the Eurasian Economic Commission dated February 12, 2016.
- The provisions of this document may be used as guidance by manufacturers, authorized representatives of manufacturers, as well as experts of the authorized bodies (expert organizations) of the Member States of the Union.
II. Concepts and Definitions
- For the purposes of applying this document, the concepts defined by the acts forming part of the law of the Union shall be used.
III. Borderline Medical Devices
- The intended purpose of a medical device is one of the main criteria for classifying a product as a medical device. The use of a medical device must provide for its medical intended purpose. Such medical intended purpose must be the sole or primary purpose.
1. Perfume and Cosmetic Products and Personal Hygiene Products
- If a product is intended by the manufacturer for use for medical purposes, such product may be classified as a medical device. Examples of products classified as medical devices:
a) instruments and devices, breast pumps, intended by the manufacturer for the treatment and prevention of diseases and pathologies of the breast;
b) hot-water bottles and heating elements, the primary purpose of which is the temporary reduction and relief of pain;
c) products (diapers, underpads, pads, etc.) for people suffering from diseases of the genitourinary system or other diseases involving impaired control of excretory functions;
d) gels-lubricants, lubricants.
e) products for laser and photoepilation.
Note. Paragraph 6 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- If a product is intended by the manufacturer for the care of the skin, hair, nails, lips, external genital organs, teeth and oral mucosa, including where the additional purpose of such product is the prevention of diseases and (or) age-related changes, such product shall not be classified as a medical device. Examples of products that are not classified as medical devices:
a) underpads, diapers, pads, etc. (except for products specified in subparagraph "c" of paragraph 6 of this document), as well as powders, dusting powders, wipes, wet wipes and other products for children and newborns;
b) bottles, nipples, rubber teething products;
c) women's personal hygiene products: pads, tampons, wipes, towels and other women's personal hygiene products;
d) personal hygiene products: antibacterial gels and liquids (sanitizers), paper tissues, wet wipes;
e) instruments, materials and products intended by the manufacturer for cosmetic procedures (permanent makeup, manicure, pedicure, piercing, epilation, etc.);
f) mattress covers;
g) heating tapes and elements, heated bottles for baby food;
h) weight-loss products (shapewear, clothing, etc.);
i) oral cavity cleaning accessories and oral hygiene products (toothbrushes for cleaning teeth and tongue, dental floss, toothpicks, toothpastes, tooth powders, teeth whitening products, mouthwashes, oral sprays, etc.);
j) general hygiene products: razors, shaving accessories;
k) products for improving appearance: wigs for women and men;
l) products for the care of the skin, hair, nails, lips, external genital organs, teeth and oral mucosa that slow down and (or) correct external signs of aging (creams, wipes, masks, body and hair washing products, decorative cosmetics, etc.);
m) antibacterial soap, gel, mousse, etc.;
n) products for intimate hygiene and care of the skin of the external genital organs: gels, foams, soaps, creams, deodorants, etc.;
o) cosmetic plasters, cosmetic eye patches;
p) products for reducing the external signs of cellulite: creams, gels, oils, scrubs, etc.
Note. Paragraph 7 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
2. Disinfectants and Equipment
- If a product is intended by the manufacturer for use for medical purposes, such product may be classified as a medical device. Examples of products classified as medical devices:
a) air recirculators intended by the manufacturer for air purification in medical organizations (except for administrative and utility premises);
b) equipment intended by the manufacturer for air disinfection in medical organizations (germicidal lamps, etc.);
c) equipment intended by the manufacturer for sterilization and disinfection of medical devices in medical organizations.
d) antiseptic wipes for treating the hands of surgeons, surgical fields and injection sites;
e) sterilization quality control indicators.
Note. Paragraph 8 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- If a product is used for treating premises, clothing and equipment for the purpose of disinfection and disinsection, as well as for air purification and maintaining a certain microclimate, including when it is used in medical organizations, such product shall not be classified as a medical device. Examples of products that are not classified as medical devices:
a) disinfectant solutions, cleaning compounds (except for products intended by the manufacturer for the special treatment of medical devices);
b) insecticidal products for treating premises against insects;
c) insect repellents (repelling agents) and acaricidal repellents for protection against insects;
d) products for combating pediculosis (shampoos, lotions, emulsion concentrates, caps or towels and wipes impregnated with a special product);
e) air recirculators, air ionizers and other air modification equipment intended by the manufacturer for household use.
Note. Paragraph 9 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
3. Products for General Purposes
- If a product is used in medical organizations for general purposes and its intended purpose does not specify use for medical purposes, such product cannot be classified as a medical device. Examples of products that are not classified as medical devices:
a) monitors, printers, scanners, telephones, fax machines, system units and other office equipment, including accessories thereto;
b) televisions;
c) refrigerators for storing products (except for equipment specified in paragraph 26 of this document).
4. Products for the Adaptation and Rehabilitation of Persons with Disabilities
- If a product is used for the adaptation and rehabilitation of persons with disabilities, such product may be classified either as a medical device or as a product for general purposes. The determining criteria are the existence of a direct relationship between the corrective action of the product and the individual needs of the patient, as well as an indication in the intended purpose of the product that it is to be used for medical purposes. Examples of products classified as medical devices:
a) accessories and prostheses for limbs;
b) hearing aids and sound amplification devices;
c) orthopedic footwear;
d) orthoses for the back and limbs;
e) walking aids for persons with disabilities;
f) crutches and canes;
g) wheelchairs;
h) lifting devices to facilitate the movement of patients (except for lifting devices specified in subparagraphs "c" and "d" of paragraph 12 of this document).
i) voice-generating devices.
Note. Paragraph 11 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- If there is no direct relationship between the corrective action of the product and the individual needs of the patient and (or) if the intended purpose of the product does not specify its use for medical purposes, such product cannot be classified as a medical device and shall be classified as a product for general purposes. Examples of products that are not classified as medical devices:
a) sound equipment for people suffering from various types of dementia, for people with visual impairments, audible pedestrian crossing signals, orientation sound signals (the equipment may be used both at home and in public places);
b) special equipment for baths and showers for persons with disabilities (chairs, seats and other devices);
c) stair lifts and lift cars for raising wheelchairs;
d) portable ramps and rehabilitation aids;
e) special ramps and lifts for baths and swimming pools for persons with disabilities;
f) devices for doors and walls for persons with disabilities;
g) tactile pedestrian crossing signals and yellow circles.
5. Products for Sports and Therapeutic Physical Training
- If the primary intended purpose of a product for sports and therapeutic physical training is its use for medical purposes in accordance with the manufacturer's intended purpose, such product may be classified as a medical device. Examples of products classified as medical devices:
a) cooling (heating) packs, plasters, bandages for pain relief;
b) cryotherapy equipment;
c) braces, bandages, dressing materials, elastic bandages and tapes intended by the manufacturer for recovery and treatment following muscle or ligament strains;
d) equipment and exercise machines intended by the manufacturer for stress tests, the primary indication for use of which is the measurement of physiological functions of the human body (with the function of measuring blood pressure, pulse and respiratory tests) (in cases where the information obtained through the use of the specified product is intended for diagnosing and making a decision on the treatment of a disease).
- Products for sports and therapeutic physical training are, in most cases, not classified as medical devices, including where their functional characteristics imply their use for medical purposes. Examples of products that are not classified as medical devices:
a) treadmills, exercise machines with a heart rate measurement function used in gyms and fitness clubs;
b) wristbands and bracelets measuring the number of steps taken and (or) heart rate;
c) bands, expanders intended by the manufacturer for muscle training and stretching.
6. Personal Protective Equipment
- If personal protective equipment is intended by the manufacturer to protect patients or medical personnel and is used for medical purposes, such personal protective equipment may be classified as medical devices. Examples of personal protective equipment classified as medical devices:
a) medical masks and medical respirators intended to limit the transmission of infectious agents;
b) surgical and examination gloves;
c) shoe covers intended by the manufacturer to prevent cross-contamination in a medical organization;
d) products for protecting patients or medical personnel against radiation, used in medical organizations during diagnostic radiology or radiation therapy;
e) mouthguards used in dental practice or for the treatment of sleep apnea;
f) protective glasses used during medical procedures;
g) clothing for medical personnel and patients used during medical procedures;
h) surgical linen (including sheets, underpads, covers used in surgical rooms);
i) soft contact lenses with zero refractive power.
Note. Paragraph 15 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- Examples of personal protective equipment that are not classified as medical devices:
a) masks and respirators for protecting the respiratory organs against aerosols, vapors and gases of hazardous and harmful substances, which are not intended by the manufacturer for use for medical purposes, including those intended for work in hazardous production conditions and emergency situations;
b) gloves and other products used, including for household purposes (cleaning, washing) (for example, shoe covers and caps used in administrative and utility premises);
c) products for protection against chemical and physical factors, including radiation, intended by the manufacturer for use in production or household settings;
d) mouthguards intended by the manufacturer for protecting teeth during sports activities;
e) products for protecting the eyes against dust, computer radiation and sunlight (protective glasses, sunglasses);
f) special clothing and special footwear that are not intended by the manufacturer for use for medical purposes.
Note. Paragraph 16 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- Excluded by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
7. Software
- Software shall be a medical device provided that it meets all of the following criteria:
a) the software is a computer program or software modules, regardless of the hardware platform used, as well as the methods of software deployment and provision of access thereto;
b) the software is not an integral part of another medical device;
c) the software is intended by the manufacturer to provide medical care;
d) the result of the software's operation consists in the interpretation of data in automatic mode, including through the use of artificial intelligence technologies, or according to parameters specified by a medical professional, which influence clinical decision-making, of a data set.
Note. Paragraph 18 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- Examples of methods and technologies used in software functions that do not constitute data interpretation:
a) displaying data obtained from a medical device, including in a specified format;
b) calculation according to specified formulas;
c) conversion between units of measurement;
d) generation of statistical reports and graphs;
e) raster or vector image editor;
f) signaling deviations in data where the original data can be displayed and provided that the deviation signaling parameters are specified by the user;
g) functions for creating screen forms, business processes, reporting and other representations used to automate the business processes of a medical organization during the operation of the software.
Note. Paragraph 19 as amended by Recommendation of the Board of the Eurasian Economic Commission dated June 29, 2021 No. 15.
- Examples of software (including examples of functions, data set sources, intended purposes, platforms, methods of providing access, etc.) that is classified as medical devices:
a) software for a physician to view an individual anatomical 3D model based on computed tomography images, meeting the following criteria:
interpretation function – calculation of the distance between two points of an anatomical 3D model;
data set source – computed tomography scanner;
intended purpose – use by radiologists, including in the provision of emergency care;
hardware platform – smartphone or tablet;
method of providing access – application store;
b) software supporting medical decision-making in stroke cases, meeting the following criteria:
interpretation function – differentiation between ischemic and hemorrhagic stroke based on diagnostic images;
data set source – computed tomography or magnetic resonance imaging scanner;
intended purpose – use by intensive care physicians, neurosurgeons and neurologists in neuro-intensive care units and vascular centers, including in the provision of emergency care;
hardware platform – any platform supporting a web browser;
method of providing access – website under the SaaS licensing model;
c) software to assist a physician in diagnosing malignant neoplasms, meeting the following criteria:
interpretation function – creation of a map of pathological changes on the skin surface, calculation of their fractal dimension to assess the degree of probability of their malignancy;
The translation preserves the structure and content of the text provided. The source text ends in the middle of subparagraph 20(c); the translation therefore ends at the same point.
Source of the dataset – healthcare professionals, by means of photographing using a smartphone's built-in camera;
Intended purpose – use by dermatovenerologists during the initial appointment;
Hardware platform – smartphone;
Method of access provision – developer's website;
g) Software to assist a physician in diagnosing tuberculous or viral meningitis in children, meeting the following criteria:
Interpretation function – analysis of cerebrospinal fluid spectroscopy data for the purpose of diagnosing tuberculous or viral meningitis in children;
Source of the dataset – spectrograph;
Intended purpose – use by a laboratory physician;
Hardware platform – personal computer;
Method of access provision – purchase of a licence and electronic storage medium;
d) Software to support medical decision-making in determining the risk of developing colorectal cancer, meeting the following criteria:
Interpretation function – assessment of the risk of developing colorectal cancer based on the available data on a patient from a high-risk group when developing an individual preventive measures plan for that patient;
Source of the dataset – healthcare professionals and diagnostic equipment;
Intended purpose – use by physicians of various specialties during initial or follow-up appointments;
Hardware platform – personal computer;
Method of access provision – purchase of a licence and electronic storage medium;
e) Software to assist a physician in diagnosing arrhythmia, meeting the following criteria:
Interpretation function – analysis of heart rhythm data for the diagnosis of arrhythmia;
Source of the dataset – pulse oximeters with a wireless data transmission interface;
Intended purpose – use by physicians of any specialty, including in the provision of emergency medical care;
Hardware platform – smartphone;
Method of access provision – application store;
f) Software to assist a physician in planning surgical techniques, meeting the following criteria:
Interpretation function – creation and visualization of an individual anatomical 3D model based on computed tomography images, used to determine the locations for placing catheters in the inner part of the bronchial tree and in lung tissue or for placing markers in soft lung tissue;
Source of the dataset – computed tomography scanner;
Intended purpose – use by surgeons (thoracic surgery, radiosurgery);
Hardware platform – personal computer;
Method of access provision – acquisition of the right to use the software with the option to download the distribution package from the developer's website;
h) Software to assist a physician in performing morphometric measurements, meeting the following criteria:
Interpretation function – image recognition and morphometry of cytological and histological specimens;
Source of the dataset – digital microscopes;
Intended purpose – use by laboratory technicians;
Hardware platform – personal computer;
Method of access provision – acquisition of the right to use the software for an unlimited period with the option to download the distribution package from the developer's website;
i) Software for remote monitoring of the health status of elderly patients with comorbid chronic diseases, meeting the following criteria:
Interpretation function – automatic detection of pathological changes in blood pressure and heart rate, signs of rhythm disturbances, followed by notification of persons providing supervision and (or) care to the patient, based on data collected and received by the central server automatically;
Source of the dataset – blood pressure monitors with a wireless data transmission interface;
Intended purpose – use by a healthcare professional;
Hardware platform – personal computer;
Method of access provision – acquisition of the right to use the software with the option to download the distribution package from the developer's website;
k) Software for developing an individual rehabilitation programme, meeting the following criteria:
Interpretation function – forecasting and assessment of the degree of the planned outcome based on the available data on the patient;
Source of the dataset – healthcare professionals and diagnostic equipment;
Intended purpose – use by rehabilitation physicians;
Hardware platform – personal computer;
Method of access provision – purchase of a licence and electronic storage medium from the developer;
l) Software used as prescribed by a physician for calculating the bolus insulin dose by a patient suffering from diabetes with a high risk of hypoglycaemia, based on carbohydrate intake, expected physical activity and blood glucose level before a meal, meeting the following criteria:
Interpretation function – selection of the prandial insulin dose;
Source of the dataset – patient and diagnostic equipment;
Intended purpose – use by the patient as prescribed by the attending physician;
Hardware platform – any platform supporting a web browser;
Method of access provision – website under the SaaS licensing model;
m) Software of a radiological picture archiving and communication system for the acquisition, storage, transmission, processing (modification of image quality, compression) and viewing of images by a physician, meeting the following criteria:
Interpretation function – determination of morphometric image parameters;
Source of the dataset – various types of diagnostic imaging equipment;
Intended purpose – use by physicians for the provision of medical care;
Hardware platform – personal computer;
Method of access provision – purchase of a licence and electronic storage medium from the developer.
Note. Clause 20 – as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
21. Examples of software that does not qualify as medical devices:
a) software intended to automate the administrative and business activities of a medical organization;
b) software, including mobile applications, intended by the manufacturer for the purposes of promoting a healthy lifestyle and developing in people a responsible attitude towards preserving and strengthening their health and maintaining active longevity, which, inter alia, measures (calculates) the number of steps, walking (running) speed, pulse, the number of calories (fluids) expended and (or) consumed, weight, body mass index, etc.;
c) medical information systems of a medical organization, laboratory information systems, software for maintaining electronic medical records, image archiving and communication systems, provided that such software does not contain data interpretation functions;
d) software, including its updates, used to control a medical device and monitor its operability;
e) software that uses data obtained from one or more medical devices but is not intended for the provision of medical care. For example, software that encrypts and (or) combines data (including patient data) obtained from one or more medical devices for their subsequent transmission;
f) software for exchanging text and (or) voice messages, electronic documents, photographic images, video and audio recordings (streams) and other data between a healthcare professional and a patient, registering, storing and providing access to such data in the provision of medical care, including through the use of telemedicine technologies, or for making appointments;
g) software for recording, planning and monitoring the implementation of scheduled maintenance and planned repairs of medical devices;
h) software intended for use by an unlimited range of users for educational, popular science and reference and information purposes, including for selecting a medical specialist, reminding users (monitoring) of the need to take a medicinal product, and providing information from the general characteristics of a medicinal product and the instructions for its medical use.
Note. Clause 21 – as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
22. Excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
23. Excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
24. Excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
25. Excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
8. Packaging and Equipment for the Storage of Medical Devices and Other Products
26. If equipment is intended by the manufacturer for the storage of medical devices and other medical products (medicinal products, etc.) requiring special storage conditions, such products may qualify as medical devices. Examples of products that qualify as medical devices:
a) equipment for the storage and transportation of biological fluids and materials, having a temperature maintenance mode and other special storage conditions;
b) equipment intended by the manufacturer for the storage of medical products, having a temperature maintenance mode and other special storage conditions;
c) packaging for the sterilization of medical devices used in medical institutions;
d) devices (injectors) for administering medicinal products supplied in replaceable cartridges.
Note. Clause 26 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
27. Packaging and equipment for the storage of medical devices and other products (medicinal products, etc.) that do not require special storage conditions do not qualify as medical devices. Examples of products that do not qualify as medical devices:
a) first-aid kit bags (first-aid kits, bags and cases for emergency medical services, cases for the storage and transportation of narcotic substances, medicinal products and devices that do not require special storage conditions, etc.);
b) eyeglass cases, contact lens cases;
c) cases for devices (injectors) for administering medicinal products supplied in replaceable cartridges.
28. The following do not qualify as medical devices:
a) primary, intermediate and secondary (consumer) packaging of medicinal products, including primary packaging of a medicinal product that serves as a means of administration (for example, a multi-dose pen injector with a permanently integrated cartridge, a pre-filled syringe);
b) a dosing device (dispenser) and (or) a device for dissolving (diluting) a medicinal product (for example, an adapter, dosing syringe, dosing cap, dosing spoon) included in the secondary (consumer) packaging of a medicinal product;
c) sachets or tablets containing a moisture absorber included in the primary or secondary (consumer) packaging of a medicinal product.
Note. Clause 28 – as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
9. Physiotherapy Equipment and Consumer Products
29. If the primary purpose of a product is physiotherapy, it is not a natural resource and is intended by the manufacturer for use only in a medical organization or as prescribed by a physician, such product may qualify as a medical device. An example of a product that qualifies as a medical device is physiotherapy equipment intended by the manufacturer for the prevention and treatment of various diseases.
30. Examples of products that do not qualify as medical devices:
a) peloids*, mineral waters, stones for stone therapy and other natural resources;
b) cedar barrels, infrared saunas and other products intended for general health improvement.
10. Furniture
31. If furniture is used in a special medical room (a treatment room, operating theatre, etc.) and (or) is subjected to a specific type of treatment, such product may qualify as a medical device. Examples of products that qualify as medical devices:
a) medical examination couches;
b) special furniture, including anaesthesiologist tables, treatment tables, infusion stands (racks), etc.;
c) massage tables;
d) operating tables;
e) functional medical beds;
f) medical chairs (dental, gynaecological, dialysis, donor chairs, etc.).
Note. Clause 31 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
32. If furniture is used outside a special medical room (a treatment room, operating theatre, etc.) and is not used for medical purposes in accordance with its primary intended purpose, such product cannot qualify as a medical device. Examples of products that do not qualify as medical devices:
a) writing desks;
b) office chairs;
c) sofas, chairs, benches and other furniture used in patient waiting rooms;
d) cabinets for storing medical products that do not require special storage conditions;
e) furniture for dining rooms in medical organizations.
11. Medical Devices Containing Medicinal Products
33. If a product contains medicinal products and its primary action is not achieved through pharmacological, immunological, genetic or metabolic action, but through physical or mechanical action, such product may qualify as a medical device. Examples of products that qualify as medical devices:
a) highly elastic materials possessing viscoelastic properties and containing hyaluronic acid and its salts, as well as other components;
b) excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15;
c) stents and other implantable devices with a drug coating, in which the antiproliferative or other pharmacological action of the medicinal product is secondary to the mechanical restoration of the lumen of a blood vessel or another part of the body achieved by means of the stent or other implantable device;
d) products for nasal and throat irrigation, including those containing substances whose primary intended action is not achieved through pharmacological, immunological, genetic or metabolic action;
e) medical devices in various forms of presentation (for example, solutions, drops, sprays, powders, pastes, gels, etc.), including those containing medicinal products, whose primary intended action is not achieved through pharmacological, immunological, genetic or metabolic action on the human body;
f) dressing materials (including wipes), adhesive plasters, wound dressings, including those containing medicinal products, antiseptic agents (including alcohols) and (or) other substances whose primary intended action is not achieved through pharmacological, immunological or metabolic action;
g) media and solutions for the transportation of organs and tissues that do not exert a metabolic effect;
h) eye drops and moisturizing solutions for eye irrigation, including those containing substances whose primary intended action is not achieved through pharmacological, immunological or metabolic action;
i) haemostatic sponges that do not contain substances involved in thrombus formation and exert a haemostatic effect by creating a mechanical barrier to bleeding;
j) concentrates and solutions for haemodialysis;
l) excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15;
m) hyaluronic acid products (for example, for intra-articular, intradermal administration, etc.);
n) excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
Note. Clause 33 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
12. Products for In Vitro Diagnostics
34. The classification of products intended by the manufacturer for laboratory use as medical devices for in vitro diagnostics shall be carried out in accordance with the definition of a medical device for in vitro diagnostics established by the acts of the Eurasian Economic Commission. Examples of products that qualify as medical devices for in vitro diagnostics:
a) haematocrit centrifuges, centrifuges for cytological examinations;
b) excluded by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15;
c) test tubes and containers for collecting human biological material;
d) plates coated with antigens or antibodies for the detection of various diseases or pathological conditions, blood groups and the Rh factor;
e) genetic tests intended by the manufacturer for diagnosing genetic disorders and diseases in humans and identifying predispositions to them;
f) culture media specifically developed and intended to provide information concerning a physiological or pathological condition using biological material obtained from a human being;
g) reagent kits intended for the extraction of nucleic acids (DNA and (or) RNA) from biological material obtained from a human being and for subsequent use in combination with a medical device for in vitro diagnostics for the purpose of detecting a specific analyte.
Note. Clause 34 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
35. If a product does not possess specific characteristics that make it possible to specially designate it for conducting medical diagnostic tests in vitro, such product cannot qualify as a medical device for in vitro diagnostics and is considered general laboratory equipment. Examples of products that do not qualify as medical devices for in vitro diagnostics:
a) reagent kits intended by the manufacturer for conducting sanitary-hygienic and sanitary-epidemiological studies, quantitative and qualitative tests of environmental objects, cosmetic, chemical and food products and food raw materials (for example, for detecting infectious disease pathogens, heavy metals, etc.);
b) reference standards, international reference materials and materials intended by the manufacturer for external quality control;
c) reagent kits intended by the manufacturer for forensic or law enforcement activities;
d) products, including equipment, measuring instruments and reagents, intended by the manufacturer for use in research and development;
e) centrifuges, pumps, mixers, scales, dispensers, incubators, microtomes, sterilizers for laboratory equipment, tissue embedding machines, microscopes, shakers, laminar flow cabinets, etc. – where the intended purpose does not specify that the products are specially intended for medical use;
f) pipettes, filters and other consumables for general purposes (disposable and reusable pipettes, plastic pipettes, Pasteur pipettes, etc.);
g) plastic and glass test tubes and vials;
h) empty enzyme-linked immunosorbent assay plates, empty Petri dishes, etc.;
i) general-purpose, non-specialized washing solutions for laboratory equipment, cell culture media, various general buffer solutions, stains, chemical reagents, etc.;
j) DNA and RNA extraction kits that provide only for the extraction of a nucleic acid sample without its mandatory intended use in combination with a medical device for in vitro diagnostics intended by the manufacturer for the detection of a specific analyte;
k) blood, plasma and serum from rabbits, calves, sheep, horses and other animals, culture media without specific medical application and other nutrient additives or auxiliary tests (for example, indole production test, oxidase test, hydrogen sulphide production test, Voges–Proskauer test).
Note. Clause 35 as amended by Recommendation No. 15 of the Board of the Eurasian Economic Commission dated 29 June 2021 No. 15.
Constitution Law Code Normative Decree Order Resolution Ruling Lawyer Almaty Lawyer Legal Services Legal Consultation Civil Criminal Administrative Cases and Disputes Legal Defence Arbitration Law Firm Kazakhstan Law Office Court Cases Declaration Decree Order Resolution Decision Report Conclusion Application Conclusion Convention Contract Memorandum Methodology Regulations Note Protocol Draft Programme Rules Messages